WELCOME TO THE NEW POCT Listserv
This site is a public meeting place that provides members of POCT Listserv with a shared calendar, discussion forums, file sharing and more. We encourage you to complete your profile and participate!
RECENT ACTIVITY
Alma Knudson posted a reply on Re: Comparability of >2 Non-Waived Devices (epocs)
Thursday, Sep 19, 2024 at 5:35 PM
Said: There are QUALITY CROSS CHECK PRODUCTS from CAP that can be use for Instrument to Instrument correlation purposes.
Sincerely,
Alma
“Far and away the best prize that life offers is the CHANCE to work hard at work worth doing” – Theodore Roos...
Leah Landis posted a new Topic
Thursday, Sep 19, 2024 at 5:12 PM
Comparability of >2 Non-Waived Devices (epocs)
I'm in the process of implementing five epocs for our respiratory department. I'm trying to come up with a procedure for our 6-month comparisons, both epoc-to-epoc and epoc-to-RapidPoint (our backup method in the main lab). I've never had more than two of the same non-waived instrument, and I'm having a hard time wrapping my mind around how to compare five identical devices to each other and to our backup method. Can I choose one epoc to be my "reference" device and compare the RapidPoin...
Danyel Olson posted a reply on Re: New CLIA changes
Thursday, Sep 19, 2024 at 4:15 PM
Said: I'm currently doing the same thing as Ivy and building my pool of "currently qualified staff" to grandfather in. I have reached out to COLA and CAP with additional questions because all the webinars seem to be focusing on all the other ...
Melissa Willcox posted a reply on Re: Lancet expiration dates for glucose meter u...
Thursday, Sep 19, 2024 at 2:09 PM
Said: At MLH in Pennsylvania, we also have found lancets with manufacture date printed on each lancet and have educated staff that the expiration date is 5 years from the manufacture date. This education is now part of the annual CBT:
Jason Ferrell posted a reply on Re: CoaguChek XS - Quality Assurance Measures?
Thursday, Sep 19, 2024 at 1:24 PM
Said: When I was using it at my old office we would send a sample for correlation to the main hospital laboratory and keep them in a record log. Maybe every 3-6 months for comparison.
MOST ACTIVE MEMBERS
James Beck
POCT Coordinator
UPMC
Peggy Mann
Ambulatory POCC & Program Manager
University of Texas Medical Branch
Kathleen David
Associate Director, Point of Care Testing
TriCore Reference Lab
Danyel Olson
POCC
Children's MN
Pet Maniquis
POCC
Providence Medical Center
Adonica Wilson
Erika Deaton-Mohney BS, MT(ASCP), CPP
POC and Compliance Coordinator
Bronson
Jeremy MacDonald
System POC Manager
Baptist Health
SHARED LINKS
No Links added yet...
UPCOMING EVENTS
This area will display upcoming events as they approach on the calendar
Add New Module