Michael Lau posted a new Topic
Monday, May 5, 2025 at 2:51 PM
Battery Impacting Abbott Freestyle Glucose Meter Integrity
Hi all. Has anyone seen the battery causing errors in the Abbott Freestyle Glucose Meter, such as an increase Strip Port errors or loss of connectivity to the wireless system?
We changed our batteries to the PROCELL INTENSE (1.5V Alkaline AA) from the Energizer Battery. There was sudden surge of Insert Strip errors, which we corrected by replacing the Strip Port. Upon a whim, I started to just replace the battery, and it resolved the issue without the change of the s...
We changed our batteries to the PROCELL INTENSE (1.5V Alkaline AA) from the Energizer Battery. There was sudden surge of Insert Strip errors, which we corrected by replacing the Strip Port. Upon a whim, I started to just replace the battery, and it resolved the issue without the change of the s...
Monika Bibbs posted a reply on Re: iSTAT cartridge traceability
Monday, May 5, 2025 at 2:40 PM
Said: Hello, I am out of office. If you require immediate assistance, please contact the main lab at 5872 for Southlake and 4247 for Northlake. Thank you.
Monika Bibbs, MHA, H(ASCP)
Manager Laboratory Services
Southlake Campus
8701 Broadw...
Stephanie Johnson posted a reply on Re: iSTAT cartridge traceability
Monday, May 5, 2025 at 2:40 PM
Said: Are your i-STATs connected to any middleware that you can see the results? You could log what you send and check for patient results within the RT timeframe. You could ask for any cartridges that were not used within that ti...
Amanda Burgess posted a new Topic
Monday, May 5, 2025 at 2:34 PM
Training vs. Competency with POC
Within the lab, we are very aware that new employees must have a training document separate from their competency verification. However, for POC, we have always had a combined sign off form for both training and verification. During our recent TJC inspection, they stated this is not sufficient and the POC operators must be treated like the lab staff and have two separate documents.
Does anyone have any advice or documents they would be willing to share to show how this can b...
Does anyone have any advice or documents they would be willing to share to show how this can b...
Amanda Burgess posted a new Topic
Monday, May 5, 2025 at 2:29 PM
iSTAT cartridge traceability
We recently participated in a TJC inspection and the inspector told us that we needed to be able to provide traceability on our iSTAT cartridges once they left storage. We currently have cartridges signed out based on quantity and labeled with RT expiration, however this is apparently not enough. He questioned how can we prove the cartridge was used within the appropriate timeframe and we should be able to trace each cartridge back to a patient.
Has anyone dealt with this or have ...
Has anyone dealt with this or have ...
MOST ACTIVE MEMBERS
James Beck
POCT Coordinator
UPMC
Peggy Mann
Ambulatory POCC & Program Manager
University of Texas Medical Branch
Kathleen David
Associate Director, Point of Care Testing
TriCore Reference Lab
Danyel Olson
POCC
Children's MN
Erika Deaton-Mohney BS, MT(ASCP), CPP
POC and Compliance Coordinator
Bronson
Pet Maniquis
POCC
Providence Medical Center
SHARED LINKS
No Links added yet...
UPCOMING EVENTS
This area will display upcoming events as they approach on the calendar
Add New Module