Said: : "Our instituttion has not defined critically ill. (Not my choice - admin's decision) We have modified the sample collection part of our Accu-Chek policy to indicate any pateint with compromised perfusion should not have a POC Accu-Chek glu...
Said: : "We are aware and nursing uses Li Heparin containers."
Said: : "These are not "critically ill" patients. They are in for an MRI scan. Even though it is considered moderately complex capillary specimens are acceptable for creatinine measurement on the NovaStat Sensor. For glucose testing in "criticall...
Said: : "Keep in mind that the Chem8 cartridge is only waived if the sample is collected in a Na or Li Heparin vacutainer."
Said: : "The i-Stat and Hemocue are considered analyzers, not meters. The fingerstick on critically ill patients is only a restriction on meters."