Said: : "We perform TAVR procedures in our CVOR and they have stated that they need that during this procedure. I believe it is due to them using the radiographic dyes during the catheterization process."
Janet is BS MT(ASCP), Point of Care Coordinator works at New Hanover Regional Medical Center
Said: : "There are also a few subtle differences between the epoc and the iStat. You cannot "blow" a cartridge with too much blood. If you overfill the epoc, the blood escapes out a vent hole and the measurements will still take place. There is ju...
Said: : "HI Pet, The current card has ionized calcium on it. No troponin or co-ox. Each card runs a 3 min calibration which checks each sensor before it allows you to add the sample. Once you add the sample it's only 40-45 seconds to get yo...
Said: : "BUN and TCO2 are now available in the EPOC. The Siemens rep had demonstrated the new cartridge.   WE were told that we may be  switching  from i-STAT to EPOC because of the cost. But we need Troponin and ionized calcium. Does anyone hav...