Said: : "We have struggled as well, our drier areas have low humidity. We have asked Siemens if they can provide guidance and perhaps indicate anti-static mats would be acceptable when humidity is low and still use the device. That was a couple mo...
Kelli is Decentralized Services Coordinator works at St. Jude Children's Research Hospital
Said: : "Yes, we do record temperature and humidity for all POCT sites.  Here in Delaware we struggle with high humidity during the summer months.  When we started to keep records, we discovered that some outpatient sites exceeded 80% RH on weeken...
Said: : "We do monitor humidity for the Clinitek. Being in New Mexico, high desert, we are more often under 18% than at or over it! We have had to put in humidifiers in several sites, which is a negative for infection control. If we are below 20%,...