Said: Clot-based tests do not need to have linearity assessed; they will not be linear. You can do a dose-response curve to prove accuracy, where increasing amounts of heparin are added to the whole blood sample and tested on the device, but ...
Said: To Erika's point though, it is not a requirement and generally not done. 
Said: ACT on which instrument? Istat, no. Medtronic HMS yes we do a linearity every 6 months per our IQCP.
Said: I can only speak to our testing and to our accreditation agency, CAP. After the initial verification study, the laboratory is not required to perform AMR verification for clot-based tests. It is best to connect with our accreditation ag...
Said: ACT is not linear. 