Point of Care Glucometers

5 followers
0 Likes

HI,


My organization currently uses Roche Accu-Chek Inform II glucometers.  I am looking at the possibility of changing to a different glucometer once our contract ends.  Any recommendations?


Stacie

35 Replies

Nova StatStrip

We also use the StatStrip and really like it.

 Just wondering if there is anyone who has switched from the Inform II glucose meter to Nova Stat Strip that can provide some feedback. 


Are the users happy with the switch?


What do the like the most?


What do they not like?


Are there any recurring problems that you are dealing with for the Stat Strip?


Are recall notices frequent?


How is Nova's technical support?


Do the contracts compare to Roche?


Is the functionality of the connectivity to RALS or your middleware the same as Roche?


Any other comments?


 


I wouldn't mind others opinions even if you haven't used Roche Inform II.  We are really considering switching over since Nova has been cleared for use in the critically ill. Smile

We made the switch about 18 months ago:


Are the users happy with the switch?  Very happy, the meters themselves do not need replacement at the rate the Roche ones did


What do the like the most?  Reliability


What do they not like? Having to accept each result, QC and patient, in order for the result to flow into LIS


Are there any recurring problems that you are dealing with for the Stat Strip?  No


Are recall notices frequent?  No


How is Nova's technical support?  Great.  On install makes Roche look laughable by comparison


Do the contracts compare to Roche?  Roche seems to write contracts with only themselves in mind.  So far so good with Nova.


Is the functionality of the connectivity to RALS or your middleware the same as Roche?  Same connectivity, but users must accept each and every result


Any other comments?  Make sure to have Nova explain all secondary costs very clearly.  We are now looking at a large bill for having to replace all battery packs (they have expiration dates on them for some reason), we were not notified of this before and at install; will now have to budget for this added cost every 2 years or so going forward.

Max we were able to set our Nova meters up so they no longer have to hit Accept, we have Telcor sending everything through and ignoring that step. We have other rules to stop glucoses from crossing, but all rules are in Telcor.

Max, you may change your flag configuration in RALS to release unaccepted results, if you wish.  Under the locations tab, flag configuration, choose device type and edit your flags.

Thank you all!

Hi Silka,


How did you setup your meters so that the user doesn't have to hit accept on the meter every time?  What rules/changes were needed in Telcor?  It sounds like something the end-user would like.


When setting it up in Telcor we had the option to allow unaccepted results to go through. You may want to call telcor I am not sure if it is something an end user can do

Sent via Groupsite Mobile.

Charlene, just give Telcor a call. I think most customers have it set up that way.

Sent via Groupsite Mobile.

I thought the accept setting is a necessary step to meet: 


GEN.43825 Result Verification Phase II


Manual and automated result entries are verified before final acceptance and reporting by the computer.


 


That is how I feel about this  “ ACCEPT” step.   If they do not  verify the Patient ID initially,


they should catch the patient ID especially if the result is critical.


I never heard our users complain about “Accept”ing  the result.


I always explain the reason why we make them jump the hoops.


Then they become liable to their actions.


I always think of this quote I have posted in my office.


“ Nothing is foolproof for  a sufficiently talented fool”. 


We should  give them every opportunity to re-check their work.


Carol Netols, there is a thing called "auto-verification" - which means the step of verification and acceptance happens automatically via a tool such as an interface manager with rules set up to detect and hold outliers for review. Most labs use this type of tool as well and are not manually verifying every single result that comes off the analyzer.

That's a good point, Silka.  Never really thought of it as autoverification, but it is.  Does that not then make the facility responsible for meeting all CAP autoverification standards?  I have always ignored those in the checklist when it comes to POC, but perhaps I shouldn't Yell...

Agree James.   The autoverification CAP requirements are a little confusing to get through.   Does anyone answer these CAP checklist items for the POC tests that go straight to the LIS? 

Displaying items 1-15 of 35 in total
Reply
Subgroup Membership is required to post Replies
Join POCT Listserv now
Stacie Kahyai
about 6 years ago
35
Replies
0
Likes
5
Followers
7783
Views
Liked By:
Suggested Posts
TopicRepliesLikesViewsParticipantsLast Reply
iSTAT in NICU -chem 8, CG4
Kim Ballister
about 14 hours ago
40125
Kim Ballister
about 8 hours ago
Rotem Sigma Validation Help
BRENDA PENA
about 16 hours ago
1085
BRENDA PENA
about 9 hours ago
Hemochron Sig Elite use outside of manufacturer temperature range
Melisa Wagoner
2 days ago
10135
Melisa Wagoner
about 17 hours ago