Said: Hello Maral,Has your transplant and perfusion team investigated through their professional circles, societies, professional organizations, this challenge they are pushing out to you/your lab team? Your venture may not be as...
Said: I think the CAP/CLIA guidelines for lab developed testing/off-label use would apply here (and would turn it into a high complexity test).  Are these samples currently tested on a different analyzer?  If so, you would want to d...
Said: We do two comparison studies.  One with our back up i-STATs to our primary i-STAT. We use 5 months of QC data to do the comparison.  We have 20 i-STATs.  We have a Primary i-STAT that all other i-STATs get compared to (we...