WELCOME TO THE POCT Listserv
This site is a public meeting place, hosted by CarePoint Solutions and moderated by Bill Donohue, that provides members of POCT Listserv with a shared calendar, discussion forums, file sharing and more. We encourage you to complete your profile and participate!
RECENT ACTIVITY
Jo Win posted a reply on Re: New Analyzer Verification of Accuracy and S...
Wednesday, Jan 14, 2026 at 12:58 PM
Said: @Jenna, I have something for epoc but the document doesn't actually talk about all the analytes that you can expand, only some. It also included this quote "These are target distributions and will rarely be completely achieved, but effo...
Jenna Ware posted a reply on Re: New Analyzer Verification of Accuracy and S...
Wednesday, Jan 14, 2026 at 12:51 PM
Said: Hi Jo! We just received a proposed validation plan from Siemens and they included ways to manipulate the samples to obtain higher or lower ranges of certain analytes. I recommend reaching out to the vendors for their recommendations.&nb...
New Analyzer Verification of Accuracy and Spanning the AMR
I have read various sources and asked CHATGPT and I am still unsure of the answer.
I have spoken to my lab director about this as well but there are conflicting opinion.
When doing a validation/verification for a new device it is preferable in an ideal world that you test patient samples that span the AMR of the device.
We do not live in an ideal world where we can get patients that span that range for the various analytes for example an analyzer like the G...
I have spoken to my lab director about this as well but there are conflicting opinion.
When doing a validation/verification for a new device it is preferable in an ideal world that you test patient samples that span the AMR of the device.
We do not live in an ideal world where we can get patients that span that range for the various analytes for example an analyzer like the G...
Chelley Schneider posted a new Topic
Wednesday, Jan 14, 2026 at 10:08 AM
Nova StatSensor Strip Creatinine Lot to Lot
Can anyone share how they perform their Nova Creatinine New Strip lot to lots?
Thank you,
Thank you,
Miranda Cotter posted a reply on Re: Gem 5000 cartridge issues
Tuesday, Jan 13, 2026 at 10:09 AM
Said: I would agree with what Cory said above - we have some time frames with many cartridge failures and then none for awhile. We have not been able to pinpoint why this occurs. We have 12 GEM5000s. I will say we do see pO2 issue...
MOST ACTIVE MEMBERS
James Beck
POCT Coordinator
UPMC
Peggy Mann
Ambulatory POCC & Program Manager
University of Texas Medical Branch
Kathleen David
Associate Director, Point of Care Testing
TriCore Reference Lab
Danyel Olson
POCC
Children's MN
Erika Deaton-Mohney BS, MT(ASCP), CPP
POC and Compliance Coordinator
Bronson
Adonica Wilson
Enterprise Point of Care Supervisor
Nemours
SHARED LINKS
No Links added yet...
Add New Module