Erika Deaton-Mohney BS, MT(ASCP), CPP posted a reply on (DELETED) Re: Blood Gas Analyzer/GEM Validation...
Tuesday, Jan 6, 2026 at 3:22 PM
Said: It looks like you have all of the information you need at your fingertips. CV% can vary based on what your mean is and what 1SD is. Some institutions use a set acceptable %CV, like 5% or so, unless stated by the manufacturer (and FDA ap...
(DELETED) Blood Gas Analyzer/GEM Validation Precision CV values and Internal vs External QC
I am trying to validate the new GEM 7000 device but this applies to the GEM 5000 devices as well.
Our Werfen rep ran external QCs for a precision study with external controls and had an observed CV of 14.3%.
Ideally shouldn't the CV be <5% ?
Should it be based on manufacturer's insert?
What is everyone's policy?
Based on the manufacturer's precision study, didn't they also fail?
Calling Werfen technical support is not very he...
Our Werfen rep ran external QCs for a precision study with external controls and had an observed CV of 14.3%.
Ideally shouldn't the CV be <5% ?
Should it be based on manufacturer's insert?
What is everyone's policy?
Based on the manufacturer's precision study, didn't they also fail?
Calling Werfen technical support is not very he...
Cory Zawikowski posted a reply on Re: Gem 5000 cartridge issues
Tuesday, Jan 6, 2026 at 8:05 AM
Said: I have 18 GEMs at my facility and have had a couple of paks kick off for similar reasons over this past weekend as well. I assign blame for PCS solution errors on the shipping process, and I had one pO2 failure and then another one that...
Carrie Leipzig posted a reply on Re: Does a PharmD qualify you as a Technical Co...
Tuesday, Jan 6, 2026 at 8:01 AM
Said: Agree with JoAnn, a credit evaluation must be completed to determine if they qualify.
Jessica Lang posted a new Topic
Tuesday, Jan 6, 2026 at 7:41 AM
Gem 5000 cartridge issues
Good morning. We have had 3 bad gem cartridges within a one week period and I was wondering if anyone else was having any issues. 2 of them leaked PCS solution in their original packaging and one was an out of the box pO2 failure which we've had before. Just a little concerned since we only have 2 instruments and this one is in our Cath Lab.
MOST ACTIVE MEMBERS
James Beck
POCT Coordinator
UPMC
Peggy Mann
Ambulatory POCC & Program Manager
University of Texas Medical Branch
Kathleen David
Associate Director, Point of Care Testing
TriCore Reference Lab
Danyel Olson
POCC
Children's MN
Erika Deaton-Mohney BS, MT(ASCP), CPP
POC and Compliance Coordinator
Bronson
Adonica Wilson
Enterprise Point of Care Supervisor
Nemours
SHARED LINKS
No Links added yet...
Add New Module