Ren Bage posted a reply on Re: COM. 04250
Wednesday, Mar 18, 2026 at 12:39 PM
Said: Erica, this is in relation to the i-STATs. Were use the QCC but not for every single i-STAT. I think it may be better to go the route of QCs! We use a mix of Coag + Chemistry + Blood Gas catridges.
Kim Ballister posted a reply on Re: COM. 04250
Wednesday, Mar 18, 2026 at 11:09 AM
Said: For those of you using iSTAT QC--same lot QC/cartridges, are you just saying that if all compared iSTATs passed QC (using eVAS) then comparison OK, or are you putting any tighter limit of acceptable comparison?
Dan Bailla posted a reply on Re: ABG Syringes (Backorder issue)
Wednesday, Mar 18, 2026 at 10:48 AM
Said: Hello,Same issue here. We got PICO 70 syringes with balanced heparin from Radiometer. This should work.
Erika Deaton-Mohney BS, MT(ASCP), CPP posted a reply on Re: COM. 04250
Wednesday, Mar 18, 2026 at 10:40 AM
Said: Hi there. This will need to revolve around what device and testing that you are speaking about. The examples for iStat are great. You can use the QC that is ran at the CLUE updates, so long as the QC level and lot number are the same. U...
Jess Reinhart posted a new Topic
Wednesday, Mar 18, 2026 at 10:25 AM
ABG Syringes (Backorder issue)
Hello!
I am not sure if other facilities are experiencing the same hardship as us, but the syringes we used for ABGs are on backorder until June 2026.... Our Respiratory team is now trying to find alternatives to utilized or alternative testing methods. So far most alternative syringes won't work because they have too many cc's of heparin and aren't big enough to gather enough blood to dilute the heparin in the syringes. As far as alternative testing methods we were thinking of ...
I am not sure if other facilities are experiencing the same hardship as us, but the syringes we used for ABGs are on backorder until June 2026.... Our Respiratory team is now trying to find alternatives to utilized or alternative testing methods. So far most alternative syringes won't work because they have too many cc's of heparin and aren't big enough to gather enough blood to dilute the heparin in the syringes. As far as alternative testing methods we were thinking of ...
MOST ACTIVE MEMBERS
James Beck
POCT Coordinator
UPMC
Peggy Mann
Ambulatory POCC & Program Manager
University of Texas Medical Branch
Kathleen David
Associate Director, Point of Care Testing
TriCore Reference Lab
Erika Deaton-Mohney BS, MT(ASCP), CPP
Technical Consultant
Sound Lab Consultant
Danyel Olson
POCC
Children's MN
Adonica Wilson
Enterprise Point of Care Supervisor
Nemours
SHARED LINKS
No Links added yet...
UPCOMING EVENTS
This area will display upcoming events as they approach on the calendar
Add New Module