Said: Same issue here for the EPOC. We have not failed any PT, but the 4% for instrument correlations seems really tight. I don't know that Siemens gives a TAE for hematocrit comparisons, but I was going to adopt the 4%. I will have to talk t...
Said: We experienced the same issue with our Avox 1000E.  As I understand it, the new CLIA 4% is for proficiency testing (PT). We kept failing PT so we discontinued the Hemoglobin on the Avox's until Werfen, the manufacturer, can address...
Said: Medicare indicated that the change that required CLIA sub-specialties occured on 10/1/2025.  I am trying to find what actually changed.  Is this somewhere on the CMS website?This is from a message I received from our r...
Said: That seems to be a counterproductive response you're receiving. The CMS Specialty Code in question is not the CLIA Subspecialties listing, try looking into this link for more information regarding this rejection: