Said: : "Sam, We do the same for new/ exchange reader as well using i-stat calver five levels calibration and two days of duplicate QC for both 1&3 and three patient comparison between lab i-stat and new/ exchange i-stat . We also used to do linea...
Said: : "I use Rals and I also like it except the fact that their tech support is not familiar with DE customization for iStat eventhough they are owned by Abbott! Sent via Groupsite Mobile."
Said: : "Thank you Sylvia. Based on what Abbott has written, they are following the CAP guidelines and considering each analyzer a testing platform. This pretty much does away with the thought/argument of the analyzer being a reader and the cartr...
Said: : "Alayna, This makes sense to me but apparently not to FDA Since it was the same reagent and test method in blue vs white cartridge of Chem8. Also any time you change reagents wheather it is the containers, vender, etc , CAP requires you to...
Said: : "I am using RALS and I like it. The only down side is the cost - per device. Are you having problems with Cobas IT?"