Said: Hello,We do the same as Jake.
Said: Hi, Our reportable range is 10 to 600 mg/dl. If the result is outside of this, it will display at >600 or <10 mg/dL and we train staff to do a venipuncture and send to the lab for specific value. 
Said: It is out of compliance to report outside of reportable range – if the result is outside of the reportable range will your course of action be different - for example (<20 or 15) or >450 or 550). We built rules to report within our reportab...
Said: I deleted my first response as I replied to the wrong issue!Invalid sample flag is likely an operator issue with scanning or entering the wrong CSN.Wrong facility - yes this happens and the CSN has to be updated to the correct C...
Said:  We’re in a similar situation. Right now our NICU team sends these samples directly to the main lab or has to draw extra blood, which we know isn’t ideal for our smallest patients. We’re also talking with Abbott about possible cart...