Said: We use the reportable range of 20-500 as that is what we were able to verify when the meters went into use. Critical Low: <40Critical High: >500 Adults, >200 Neonates
Said: I have the instructions that they gave me when they installed our Hemochron SE instruments.
Said: We performed linearity studies when first implemented (2018) and we confirmed our AMR is 10-600 mg/dL. We don't perform linearity studies on every new meter. We complete performance verification on every new meter and perform interface ...
Said: Ask you company (Hemochron) of their procedure. It is can be found in the instrument operator’s manual. Sincerely, Alma “Far and away the best prize that life offers is the CHANCE to work hard at work worth doing” – Theodore Roosevelt Alm...
Said: Our medical director had us perform a study with 20 Nova StatStrip meters and 50 "patients". Some samples were left on the table overnight (or longer) and some were spiked. We took the highest low and the lowest high as the AMR. All of the...