Alma Knudson posted a reply on Re: Glucose Meter Reportable Ranges
Thursday, Aug 14, 2025 at 11:46 AM
Said: Per CLIA and other regulatory agency – reportable range (is also known as Linearity or AMR (Analytical Measurement Range) is based on the result that your meter can recover using a KNOWN material like LINEARITY MATERIAL that your meter compa...
Adelina Wright posted a reply on Re: Cholestech LDX and RALS
Thursday, Aug 14, 2025 at 11:23 AM
Said: The instrument does not allow patient or operator information to be entered at the time of the test, so you would need to enter this info once the result hits RALS and then manually push it for it to transmit to the LIS. We decided not ...
Adelina Wright posted a new Topic
Thursday, Aug 14, 2025 at 11:14 AM
Glucose Meter Reportable Ranges
We currently use the Nova StatStrip Hospital Meter at our instituation. The reportable range that we have been using for 10+ years for our adult meters is 60 - 450 mg/dL even though the meter can technically measure from 10 - 600 mg/dL. This aligns with the lower critical value that we have set as less than 60. Recently we have had units request that their meters measure lower than 60 mg/dL and we are curious if our ranges align with other institutions.
What reportable range do yo...
What reportable range do yo...
Sonya Sheppard posted a reply on Re: Siemens epoc QC Monitoring
Wednesday, Aug 13, 2025 at 2:51 PM
Said: For the Epoc, we monitor card rejection rate with a threshold of 10%. When we are above this, most of the time we see it is user errors and provide an inservice and rates come back in range. There was once when we had to get...
Gina Smith posted a new Topic
Wednesday, Aug 13, 2025 at 2:31 PM
Siemens epoc QC Monitoring
Good afternoon,
Our institution chose to monitor QC errors on the Siemens epoc as a quality metric. We based our acceptable limits on historical Abbott iSTAT data, which the epoc replaced.
Unfortunately, we are finding that our end users are not yet able to meet this metric. We are curious to hear if others are doing the same, and if so what is your threshold, or is anyone perhaps monitoring another aspect of this test method.
Appreciate your valued ins...
Our institution chose to monitor QC errors on the Siemens epoc as a quality metric. We based our acceptable limits on historical Abbott iSTAT data, which the epoc replaced.
Unfortunately, we are finding that our end users are not yet able to meet this metric. We are curious to hear if others are doing the same, and if so what is your threshold, or is anyone perhaps monitoring another aspect of this test method.
Appreciate your valued ins...
MOST ACTIVE MEMBERS
James Beck
POCT Coordinator
UPMC
Peggy Mann
Ambulatory POCC & Program Manager
University of Texas Medical Branch
Kathleen David
Associate Director, Point of Care Testing
TriCore Reference Lab
Erika Deaton-Mohney BS, MT(ASCP), CPP
Technical Consultant
Sound Lab Consultant
Danyel Olson
POCC
Children's MN
Adonica Wilson
Enterprise Point of Care Supervisor
Nemours
SHARED LINKS
No Links added yet...
UPCOMING EVENTS
This area will display upcoming events as they approach on the calendar
Add New Module