WELCOME TO THE POCT Listserv
This site is a public meeting place, hosted by CarePoint Solutions and moderated by Bill Donohue, that provides members of POCT Listserv with a shared calendar, discussion forums, file sharing and more. We encourage you to complete your profile and participate!
RECENT ACTIVITY
Tonya Sabin posted a reply on Re: JACO Officially listed i-STATs as a Blood A...
Tuesday, Sep 8, 2026 at 1:47 PM
Said: Hello all,I just wanted to circle back to this discussion. In April we found out about this change (our JC inspection was in Feb), then found out CMS was coming the following week for review. We wanted to get ahead o...
Kimberly Arnesen posted a reply on Re: Siemens DCA Replacement Validation Protocol
Tuesday, Sep 8, 2026 at 1:15 PM
Said: Hi,for new or replacement waived test instruments we have a validation summary page we use for all our devices with device info, serial number, date device was checked and tagged by Biomed, and that initial QC was performed and pass...
Rachel Donner posted a new Topic
Tuesday, Sep 8, 2026 at 10:08 AM
Siemens DCA Replacement Validation Protocol
We need to set up a replacement DCA for one of our outpatient clinics. DCA HA1cs as a new platform were initially validated over 10 years ago and included extensive validation testing (accuracy, precision, comparison, linearity). Given that the DCA HA1c is a waived test and we are regulated by CLIA/CAP, do you have any recommendations for replacement instrument validation protocols? We contacted the manufacturer and they said to follow our internal regulatory protocol (which would be...u...
Anil Chokkalla posted a reply on Re: hs-Tnl verification- precision criteria
Monday, Sep 7, 2026 at 11:16 PM
Said: To meet the 4th universal definition of "high sensitivity" cardiac troponin assay, I would utilize <=10% CV at the 99th percentile upper reference limit. This limit could also be applied for i-STAT hs-TnI Controls Levels 1-3....
Mandye Canales posted a new Topic
Monday, Sep 7, 2026 at 3:24 PM
hs-Tnl verification- precision criteria
Hello,
We're working on our verification of the hs-Tnl iSTAT cartridges. I am looking for assistance on acceptable criteria for precision for this analyte. After reviewing the IFU and method evaluation from Abbott, I cannot find criteria with supporting evidence. The precision in the IFU is not to be used for precision criteria per the method evaluation. The method evaluation states that precision must include SD and %CV but doesn't provide guidance.
Does anyone who has...
We're working on our verification of the hs-Tnl iSTAT cartridges. I am looking for assistance on acceptable criteria for precision for this analyte. After reviewing the IFU and method evaluation from Abbott, I cannot find criteria with supporting evidence. The precision in the IFU is not to be used for precision criteria per the method evaluation. The method evaluation states that precision must include SD and %CV but doesn't provide guidance.
Does anyone who has...
MOST ACTIVE MEMBERS
James Beck
POCT Coordinator
UPMC
Peggy Mann
Ambulatory POCC & Program Manager
University of Texas Medical Branch
Kathleen David
Associate Director, Point of Care Testing
TriCore Reference Lab
Erika Deaton-Mohney BS, MT(ASCP), CPP
Technical Consultant
Sound Lab Consultant
Danyel Olson
POCC
Children's MN
Adonica Wilson
Enterprise Point of Care Supervisor
Nemours
SHARED LINKS
No Links added yet...
UPCOMING EVENTS
This area will display upcoming events as they approach on the calendar
Add New Module