Christin Maurer posted a reply on Re: iSTAT Cartridge lot to lot
Friday, Jul 24, 2026 at 2:53 PM
Said: I can definitely appreciate that perspective. However, the guidance we've received is that CMS now considers the i-STAT an analyzer rather than a reader, and TJC has communicated that expectation to accredited laboratories. As a result,...
Tracy Learned posted a reply on Re: iSTAT Cartridge lot to lot
Friday, Jul 24, 2026 at 2:32 PM
Said: I have been told by Abbott tech support that the i-STAT is NOT an analyzer, despite the info circulating otherwise. Also, my interpretation of COM.30450 is that lot to lot comparison is not required. It reads "New reagent lots and shipm...
Christin Maurer posted a reply on Re: iSTAT Cartridge lot to lot
Friday, Jul 24, 2026 at 12:08 PM
Said: For istat to istat within the same dept, we are utilizing QC for CG8+ and ACT-C (a 6-month look back). For dept to dept (including main lab), we use one istat per dept and then use our main lab blood gas to complete blood gas patient co...
Erin Walters posted a new Topic
Friday, Jul 24, 2026 at 11:19 AM
iSTAT Cartridge lot to lot
Hello Everyone!
I wanted to see if anyone will share what y'all are doing with the new cartridge lot to lot. Now that the iSTATs are changed to an analyzer and QC needs to be performed on each analyzer, I want to confirm my thoughts that we need to do the lot to lot on every iSTAT also?
Also, with ACT correlations is anyone designating a primary iSTAT and what materials are you using for the ACT correlations?
Thanks so much!
Erin
I wanted to see if anyone will share what y'all are doing with the new cartridge lot to lot. Now that the iSTATs are changed to an analyzer and QC needs to be performed on each analyzer, I want to confirm my thoughts that we need to do the lot to lot on every iSTAT also?
Also, with ACT correlations is anyone designating a primary iSTAT and what materials are you using for the ACT correlations?
Thanks so much!
Erin
Jeanne Mumford posted a reply on Re: POCC Mixer Event in Anaheim!
Friday, Jul 24, 2026 at 9:51 AM
Said: Bill,You are so gracious! I will send my RSVP today and I appreciate the opportunity to mingle with colleagues Monday evening. See you all soon.
MOST ACTIVE MEMBERS
James Beck
POCT Coordinator
UPMC
Peggy Mann
Ambulatory POCC & Program Manager
University of Texas Medical Branch
Kathleen David
Associate Director, Point of Care Testing
TriCore Reference Lab
Erika Deaton-Mohney BS, MT(ASCP), CPP
Technical Consultant
Sound Lab Consultant
Danyel Olson
POCC
Children's MN
Adonica Wilson
Enterprise Point of Care Supervisor
Nemours
SHARED LINKS
No Links added yet...
UPCOMING EVENTS
This area will display upcoming events as they approach on the calendar
Add New Module