Siemens Multistix 10 SG
1 followers
0 Likes
Anyone seeing an increase in false positive (Trace-Intact) blood results using Multistix 10 SG lot number 704054? We use them on the Clinitek Status+ and have seen the number of failures on our negative QC quadruple and have also had an unacceptable proficiency result.
1 Reply
Reply
Subgroup Membership is required to post Replies
Join POCT Listserv now
Suggested Posts
| Topic | Replies | Likes | Views | Participants | Last Reply |
|---|---|---|---|---|---|
| Labeling of POCT samples and devices | 1 | 0 | 158 | ||
| LeadCare II sample collection | 1 | 0 | 130 | ||
| GH100 Training | 1 | 0 | 153 |
We do not currently have that lot number. I have noticed however, depending on the control manufacturer, Negative-Trace is an acceptable range on Blood for Level 1/Negative.