0 Likes

Anyone seeing an increase in false positive (Trace-Intact) blood results using Multistix 10 SG lot number 704054? We use them on the Clinitek Status+ and have seen the number of failures on our negative QC quadruple and have also had an unacceptable proficiency result.

1 Reply

We do not currently have that lot number. I have noticed however, depending on the control manufacturer, Negative-Trace is an acceptable range on Blood for Level 1/Negative.

Reply