I am so glad Listserv is back up and running!


For those that are using STATNotes and Meditech, do you know what the result codes are in the OBX 3 segment?  I am trying to get it started at our facility and I have it built and it is being received in Telcor, but I don't know what code Telcor needs to send it to Meditech.


 


Thanks!

2 Replies

Hi Jillian, it is the exact way the specific test is built in Meditech. For example, for arterial pH our Meditech result code that our IT LIS team built in the Meditech disctionary is ph-art


So, Telcor needs to know to use lowercase ph-art in OBX3. When you are building statnotes it is the same thing, each parameter is built in Meditech as a test in the dictionary, such as delsys for delivery system - so Telcor needs to know the exact result code built in the Meditech dictionary for each of your new Statnotes parameters.


Here is a sample MSH from our Meditech, I entered X's for the patient name and ID:


 



MSH|^~\&|34|MGA|||201804101737||ORU^R01|MSG201804104E4C7ED0|P|2.3|||AL|AL||
PID|1||||XXXXX^XXXXX^^||XXXXXXXXX|M||||||||||208755288
ORC|NW||
OBR|1|||AG|||201804101734|||111855||||201804101735|ARTERIAL||||||||||F|||||PORT
|||111855


OBX|1|ST|ph-art||7.351||||||F|||201804101734||111855


OBX|2|ST|pco2-art||42.4|mmHg|||||F|||201804101734||111855


OBX|3|ST|po2-art||58|mmHg|||||F|||201804101734||111855
OBX|4|ST|tco2-art||25|mmol/L|||||F|||201804101734||111855


OBX|5|ST|be-art||-2|mmol/L|||||F|||201804101734||111855


OBX|6|ST|hco3-art||23.7|mmol/L|||||F|||201804101734||111855


OBX|7|ST|o2sat-art||90|%|||||F|||201804101734||111855


OBX|8|ST|oper id-poc||111855||||||F|||201804101734||111855


OBX|9|ST|site-poc||380947||||||F|||201804101734||111855


OBX|10|ST|pt-temp||35.9||||||F|||201804101734||111855


OBX|11|ST|istat-com 1||NP||||||F|||201804101734||111855


OBX|12|ST|istat-com||NP||||||F|||201804101734||111855

Thanks!!!

Reply
Subgroup Membership is required to post Replies
Join POCT Listserv now
Jillian Cooper
over 6 years ago
2
Replies
0
Likes
0
Followers
746
Views
Liked By:
Suggested Posts
TopicRepliesLikesViewsParticipantsLast Reply
Hemochron Sig Elite use outside of manufacturer temperature range
Melisa Wagoner
1 day ago
1091
Melisa Wagoner
28 minutes ago
EPOC scanner issue post software update
Ivy Douglas
5 days ago
60307
Ivy Douglas
4 days ago
PFA-100
Autilia Sisti
6 days ago
00121
Autilia Sisti