Failed EPOC QC

0 followers
0 Likes

Hi,


I'm having issues with failing EPOC QC. QC is Eurotrol, GAS-ISE metabolites level 2.


I've tried 2 different lot numbers of QC, 266-2-B837 and 266-2-B809, two different lots of cards and tried different EPOC instruments.


All level 2 QC has failed on all devices. Levels 1 and 3 pass, but they are at the low end of the acceptable QC range.


Is anyone else having these issues?


Thanks,


Jake

1 Reply

Jake,


We have lot# 266-2-B809 and the controls look fine. I would maybe confirm that your value assignments or eVAS are up to date. 


Karen


 