Failed EPOC QC

0 followers
0 Likes

Hi,


I'm having issues with failing EPOC QC. QC is Eurotrol, GAS-ISE metabolites level 2.


I've tried 2 different lot numbers of QC, 266-2-B837 and 266-2-B809, two different lots of cards and tried different EPOC instruments.


All level 2 QC has failed on all devices. Levels 1 and 3 pass, but they are at the low end of the acceptable QC range.


Is anyone else having these issues?


Thanks,


Jake

1 Reply

Jake,


We have lot# 266-2-B809 and the controls look fine. I would maybe confirm that your value assignments or eVAS are up to date. 


Karen


 

Reply
Subgroup Membership is required to post Replies
Join POCT Listserv now
Jake Fray
almost 7 years ago
1
Reply
0
Likes
0
Followers
600
Views
Liked By:
Suggested Posts
TopicRepliesLikesViewsParticipantsLast Reply
Does a PharmD qualify you as a Technical Consultant
McKenna Chandler
3 days ago
00267
McKenna Chandler
3 days ago
Upgrading Middleware Validations Plan (RALS)
Kristy Saunders
5 days ago
41383
Amanda Casavant
4 days ago
Humidity Monitoring
Jennifer Gagne
7 days ago
100573
Kathleen David
5 days ago