We are validating a new LIS system and need to make sure that results that fall out of the useable range flag properly in the LIS (i.e. <10 or >600).  High flags are easy for all tests.  However I need to know what to do to get a < flag for ACT and PT/INR?


What can I do to the specimen to make it come out that low?

2 Replies

Have you tried using partially activated blood? 

Hi Bob,


     Have you tried dosing the cartridge and not hitting the start button for a while? Seems to work for the operators... just sayin'.  :)


lois.snider@stclair.org


 