Rapid Point 500 hypoxic AMR verification

1 followers
0 Likes

Hello!  What is everyone using for the very low pO2 AMR verification on their RapidPoints?  There has been a concerned raised for the Eurotrol since they don't provide a target value.


Thanks!


Danielle

3 Replies


We use the Siemens’ Cal Ver for RP500 but it is not Eurotrol.


Manufacturers description: Calibration verification material linearity blood gas Rapidpoint 400. (We run this on our 500s and have good results.)









This message may contain confidential information protected by law through attorney-client privilege or professional peer review/quality evaluation privilege. It is intended only for the individual or entity named above. It is prohibited for anyone else to
disclose copy, distribute, or use the contents of this message. If you are not the intended recipient, please contact the sender by reply e-mail and destroy all copies of the original message. You may also contact the Bronson Healthcare Group Privacy Officer
at privacyofficer@bronsonhg.org

How low does the PO2 go on the Siemens CalVer?  We are hoping to verify down to 15 mmHg for our cord bloods.


Thanks!

I just completed the most recent cal ver. PO2 low is only 27 mmHg.

Reply
Subgroup Membership is required to post Replies
Join POCT Listserv now
Danielle Ennis
over 5 years ago
3
Replies
0
Likes
1
Followers
1197
Views
Liked By:
Suggested Posts
TopicRepliesLikesViewsParticipantsLast Reply
Roche Chemstrip specific gravity
Laura Ball
4 days ago
10113
Michael Bishop
4 days ago
Anybody have experience with the Actalyke?
Edith Synnefakis
5 days ago
00148
Edith Synnefakis
5 days ago
Cal/Ver Istat pCO2
Autilia Sisti
5 days ago
20269
Autilia Sisti
4 days ago