Hello,


Does anyone have their Medtronic ACT Plus devices interfaced?  I have some questions about the operator workflow that Medtronic seems unable to answer.


Can testing personnel accept or reject a result?


Thanks!


Danielle

1 Reply

Our Medtronic HMS Plus's are interfaced. We have Rals and Epic. I had IT build it where the results stopped in the flagged results in Rals (in case of data entry error by operator). I review and send to Epic. I can also edit in the operators ID once in Epic.


No, testing personnel cannot accept or reject a result.


 


Kenda Webb, MT(ASCP)


Supervisor, POC


Laboratory