i-STAT Validation
20 followers
0 Likes
We are updating some of our i-STAT fleet to align with current wireless security standards, 5GHz. Curious if you all are validating the cartridge or the analyte. Thanks!
6 Replies
Reply
Subgroup Membership is required to post Replies
Join POCT Listserv now
Suggested Posts
| Topic | Replies | Likes | Views | Participants | Last Reply |
|---|---|---|---|---|---|
| How many full-time employees does your POCT department staff? | 10 | 0 | 567 | ||
| Hemachron 100 vs. iSTAT and HMS | 0 | 0 | 284 | ||
| AMR verification of hemoglobin derivatives | 4 | 0 | 328 |
We opened a ticket and created a project with both the vendor (Abbott) and our IT team and LIS. We tested the functionality and made sure that it worked by looking for results for cartridges (CG8, Chem8, ACT....) in use in the EMR, without having to dock the meter.
Hope this helps.
Angel Lorenzo