WELCOME TO THE POCT Listserv
This site is a public meeting place that provides members of POCT Listserv with a shared calendar, discussion forums, file sharing and more. We encourage you to complete your profile and participate!
RECENT ACTIVITY
Monika Bibbs posted a new Topic
Thursday, Mar 21, 2024 at 4:47 PM
Comparisons Abbott i-STAT vs IL GEM 5000
Hello,
I'm in search of suitable material to perform comparisons of ionized calcium on the Abbott i-STAT (EG7+ blood gas cartridge) and the Instrument Laboratories GEM Premier 5000. CAP has samples that are specific to the i-STAT but not the GEM, and vice versa. Anyone have any suggestions.
I'm in search of suitable material to perform comparisons of ionized calcium on the Abbott i-STAT (EG7+ blood gas cartridge) and the Instrument Laboratories GEM Premier 5000. CAP has samples that are specific to the i-STAT but not the GEM, and vice versa. Anyone have any suggestions.
Monica Kaufman posted a new Topic
Thursday, Mar 21, 2024 at 4:18 PM
Coaguchek Pro New Strip Correlation
We have a new lot of testing strips, so are running a patient correlation prior to putting in place in our AntiCoag Clinic. For our correlation, we run the current lot, the new lot, and send a sample to the main lab to run on their analyzer. Our new lot #73389513 is running higher than the current lot and the lab instrument. Variation is anywhere from 0.4 - 0.8. Has anyone else seen or had this issue? I have a call placed into Roche, but this group is the be...
Rosemary Gray posted a reply on Re: ACT correlation_ Hemochron vs istat
Thursday, Mar 21, 2024 at 11:05 AM
Said: That study sounds interesting and a great idea. Hopefully you can share your data with us, I would appreciate it. Thank you for sharing.
Christina Schleich posted a reply on Re: ACT correlation_ Hemochron vs istat
Thursday, Mar 21, 2024 at 10:53 AM
Said: Thank you Silka for the great reply!We are also currently having this issue in our Cathlab thinking they are using more heparin lately with the i-Stat ACT-k than when they had Hemochron LR. We are actively doing a 30 patient c...
Adonica Wilson posted a reply on Re: ACT Plus Liquid QC
Thursday, Mar 21, 2024 at 9:29 AM
Said: The other thing to look at is your state requirement. Some states have different QC requirements, so just check them to make sure the frequency isn't required more than the manufacturer.
MOST ACTIVE MEMBERS
James Beck
POCT Coordinator
UPMC
Peggy Mann
Ambulatory POCC & Program Manager
University of Texas Medical Branch
Kathleen David
Associate Director, Point of Care Testing
TriCore Reference Lab
Danyel Olson
POCC
Children's MN
Erika Deaton-Mohney BS, MT(ASCP), CPP
POC and Compliance Coordinator
Bronson
Pet Maniquis
POCC
Providence Medical Center
SHARED LINKS
No Links added yet...
UPCOMING EVENTS
This area will display upcoming events as they approach on the calendar
Add New Module