WELCOME TO THE POCT Listserv

This site is a public meeting place, hosted by CarePoint Solutions and moderated by Bill Donohue, that provides members of POCT Listserv with a shared calendar, discussion forums, file sharing and more. We encourage you to complete your profile and participate!

THIS GROUP CONSISTS OF

Group Members
You can connect with
Member Objectives
You can assist with
Files
You can download

RECENT ACTIVITY

Said: My facility has been encountering an issue in which patients admitted from the ED are being "discharged" by our middleware. This creates a problem at bedside when operators scan the patients admission wristband. The CSN account has been...
Said: I cannot state how relieved I am that this does not happen at my facility.  We do have plastic arm bands, and we use a 12 digit encounter number, not the medical record (whose digits vary).  By only allowing 12 digits on the g...
Said: This issue has been a huge thorn in my side!!! I agree if people would please check the scan is correct when the meter says it isn't before selecting override a lot of this wouldn't be an issue.  But I dream and digress.Sta...
Said: Hi, we have this statement in our Analytical Correlation workbook and procedure, approved by our Senior Clinical Laboratory Director. " Results must agree within the Total Allowable Error limits for like devices. *** For i-STAT to Sysme...
Said: We experienced this as well.  We could hit 4% for the Instrument Correlations but not Method.  I had searched and read on this list serv that the 4% was acceptability criteria for proficiency testing.  I created an SBAR f...