WELCOME TO THE POCT Listserv

This site is a public meeting place, hosted by CarePoint Solutions and moderated by Bill Donohue, that provides members of POCT Listserv with a shared calendar, discussion forums, file sharing and more. We encourage you to complete your profile and participate!

THIS GROUP CONSISTS OF

Group Members
You can connect with
Member Objectives
You can assist with
Files
You can download

RECENT ACTIVITY

Said: For the most part, we follow main laboratory ranges for the Chemistries and the Blood gas ranges for the normals and critical values that RT and the pulmonologists have approved. This means that we do not have to do normal rang...
Said: In our system, all test panels that contain failed tests are held in the POC middleware.  The users are instructed to either rerun the test OR they can submit a POC Correction form with the request to push the remaining tests to th...
Said: Not only does CAP require an IQCP, but CLIA (Joint Commission) does as well.  CLIA states that for a moderately complex analyzer that does not run the minimal CLIA QC requirement (2 levels of external controls once a day) then they...
Said: DanielleWe have 9 TEG S at our facility that we use for POC only.    We track our QC through the TEG Manager which provides dates and lot numbers. Thanks, Dan