Andries Sigtenhorst posted a reply on Re: Abaxis Piccolo
Wednesday, Feb 18, 2026 at 1:17 PM
Said: The Sodium fails because you are outside the measurable range and your "~~~" indicates that. Try running in QC mode which has an "extended" reportable range. We encountered the phenomenon with BRT and NovaOne controls for Level 1 Sodium...
Ashlee Byquist posted a new Topic
Wednesday, Feb 18, 2026 at 12:34 PM
Abaxis Piccolo
Does any one use the Abaxis Piccolo and seem to have trouble with the level one control not reporting out the Sodium? It seems we have this same issue every month that we run it. We try to open new bottles, get a new lot of QC, and sometimes order new QC from company. We can get it to come in most of the time after a lot of troubleshooting and repeats, but I just don't understand why we have so much trouble with it every time. Level 2 always seems to pass just fine. All other parameters ...
Amanda Casavant posted a reply on Re: Reference Ranges - ABL90
Wednesday, Feb 18, 2026 at 12:03 PM
Said: Hi HeatherIt is really difficult to validate blood gas reference ranges. When our organization switched blood gas analyzers, we looked at the "new" ranges (ABG and VBG) vs our established ranges and did a literature review. There wa...
Heather Matalka posted a new Topic
Wednesday, Feb 18, 2026 at 11:21 AM
Reference Ranges - ABL90
Hi all,
For other hospitals with Radiometer's ABL90, how do you handle reference ranges (finding literature and in house verification) for more difficult sample types such as capillary for oximetry values and pO2 for venous blood?
For other hospitals with Radiometer's ABL90, how do you handle reference ranges (finding literature and in house verification) for more difficult sample types such as capillary for oximetry values and pO2 for venous blood?
Heather Matalka posted a reply on Re: HMS+ ACT >999 reporting
Wednesday, Feb 18, 2026 at 11:19 AM
Said: Hi Angie, We have the same problem! We also have RALS and whenever the HR ACT + HPT are run together, if the HR-ACT reads as >999, we don't get an HR-ACT result at all. We currently don't have a way to get the HPT result...
MOST ACTIVE MEMBERS
James Beck
POCT Coordinator
UPMC
Peggy Mann
Ambulatory POCC & Program Manager
University of Texas Medical Branch
Kathleen David
Associate Director, Point of Care Testing
TriCore Reference Lab
Erika Deaton-Mohney BS, MT(ASCP), CPP
Technical Consultant
Sound Lab Consultant
Danyel Olson
POCC
Children's MN
Adonica Wilson
Enterprise Point of Care Supervisor
Nemours
SHARED LINKS
No Links added yet...
UPCOMING EVENTS
This area will display upcoming events as they approach on the calendar
Add New Module