WELCOME TO THE POCT Listserv

This site is a public meeting place, hosted by CarePoint Solutions and moderated by Bill Donohue, that provides members of POCT Listserv with a shared calendar, discussion forums, file sharing and more. We encourage you to complete your profile and participate!

THIS GROUP CONSISTS OF

Group Members
You can connect with
Member Objectives
You can assist with
Files
You can download

RECENT ACTIVITY

Said: @Don Tran, I would not expand past the reportable range of the manufacturer's recommendation. I would use the internal QC data (which they say are equivalent to external liquid qc) from the machine to verify the manufacture...
Said: When you expand the reportable range beyond the manufacturer's recommendation, aren't you deviating from it? In that case, wouldn't it change the test complexity? Would non-lab personnel still be qualified to perform it? I would verify ...
Said: I interpret method specific target values to mean the control material package insert specifically states a range for the GEMS, or a range for i-STAT, or a range for ABL analyzers, etc.  If the control material you are using is spe...
Said: Jamey, so many times in our 'one POC Program' program, I've seen us come up with solutions which were not in the least compromises from 'several sides' of the issue. Is it possible to ignore those who believe 'if they get to use fake te...
Said: @Lori Thank you so much. You are amazing for calling CAP, they are very helpful just like you. so if Werfen GEM manual says the internal solution controls (PCSs) are "traceable to NIST Primary standards...Th...

UPCOMING EVENTS

This area will display upcoming events as they approach on the calendar