Said: We go ahead and correlate both HCT & HGB between the XN and Epoc.
Said: Yes, there is a difference. My response is if they are TRUELY where one is calculated and one is measured. My scenario was where they are not like for like. Jim shared details that he shows : The HCT is considered a directly...
Said: Thank you Erika and Jim, for your answers. To clarify, this is for the semi annual method correlations. There is a difference in Erika and Jim's response. Erika: ".you don't need to do any further co...
Said: hi Joon the XN analyzer, the HCT is considered a directly measured analyte by cumulative pulse height of the RBCs as they pass through the RBC aperture. It should be included in the correlation studies. The HB is also directly measu...
Said: I love this question!! You are correct that you are not "required" to method compare. Are you speaking in terms of initial verification or twice per year correlations?  In the Verification Studies, a calculation from one method to ...