HEMACHRON SIGNATURE ELITE CUVETTE USE IN CATH LAB
11 followers
0 Likes
We are transitioning from the Hemochron response to the Hemochron Signature Elite. For most cases the LR cuvette shows good correlation with the response. Some cases require higher ACT results so the Cath Lab is getting > results with the LR cartridge. The docs want a number and not just a >. Does anyone use both the LR cartridges and the + cartridges in their Cath Lab settings?
Rick Deerfield
POCC
Penn State Health St. Joseph
Rick Deerfield
POCC
Penn State Health St. Joseph
6 Replies
Reply
Subgroup Membership is required to post Replies
Join POCT Listserv now
Suggested Posts
Topic | Replies | Likes | Views | Participants | Last Reply |
---|---|---|---|---|---|
iSTAT in NICU -chem 8, CG4 | 6 | 0 | 268 | ||
Rotem Sigma Validation Help | 1 | 0 | 106 | ||
Hemochron Sig Elite use outside of manufacturer temperature range | 2 | 0 | 176 |
We are actually in the process of validating the ACT + cartridge for those Cath Lab patients where there is a higher ACT expected.
We also use both LR and + Cartridges - but never interchangeably. If there is an anticipated high ACT, start with the + and stay with it. Our perfusion team uses ACT+ almost exclusively. We were always instructed that the results in the <400 range may not have good agreement between cartridge types. Michele Naples