Validation of replacement Accu-Chek Inform II meters
7 followers
0 Likes
Would anyone be able to share their process for validation of replacement Accu-Chek Inform II meters?
One of our floors had a rough week and broke two of them.
Thank you,
Tracey
One of our floors had a rough week and broke two of them.
Thank you,
Tracey
6 Replies
Reply
Subgroup Membership is required to post Replies
Join POCT Listserv now
Suggested Posts
| Topic | Replies | Likes | Views | Participants | Last Reply |
|---|---|---|---|---|---|
| Does a PharmD qualify you as a Technical Consultant | 0 | 0 | 172 | ||
| Upgrading Middleware Validations Plan (RALS) | 4 | 1 | 357 | ||
| Humidity Monitoring | 10 | 0 | 556 |
We have NOVA Stat meters (CLIA waived).
Laboratory Director's approved validation plan:
~Perform a linearity (5 Levels, minimum of 2 replicates) and both levels of QC.
This seems to be the standard practice for many hospital based glucometers.