Hemochron ACT Cuvette Shipment/Lot Verification
9 followers
0 Likes
Hello everyone,
For anyone who also uses Hemochron cuvettes, how do you perform lot-lot testing? Do you calculate acceptable variance?
Thank you in advance!
For anyone who also uses Hemochron cuvettes, how do you perform lot-lot testing? Do you calculate acceptable variance?
Thank you in advance!
4 Replies
Reply
Subgroup Membership is required to post Replies
Join POCT Listserv now
Suggested Posts
| Topic | Replies | Likes | Views | Participants | Last Reply |
|---|---|---|---|---|---|
| Roche Chemstrip specific gravity | 1 | 0 | 107 | ||
| Anybody have experience with the Actalyke? | 0 | 0 | 141 | ||
| Cal/Ver Istat pCO2 | 2 | 0 | 262 |
What are others using?
Lot to lot comparison is accomplished in the following manner:
Example, ACT-LR: (156 - 163) ÷ 163 x 100 = 4.3% difference
*ACT-LR, ACT+ and PT-INR are all clot based tests. Our laboratory uses Evaluation Criteria for INR as published in the most recent CAP Participant Summary for WP6.
Also are your instruments interfaced?