Hello all,

Who here uses the Triage MeterPro Cardiac Panel for Troponin I and doesn't mind answering some questions I have? 

5 Replies

We use it as a backup method

Great! Thank you for responding, Tara. We are trying switch from i-STAT to Triage for Troponin I and having problems with patient correlations for "abnormal" patient results. (The "normal" results are correlating). We have compared Triage to i-STAT and Triage to Vitros, and still having the same issues and not able to get within 30% difference. CAP PCARM surveys are acceptable, but results differ still from one another due to matrix effect. There are multiple factors in play that we have tried to consider and have exhausted all options so far that we can think of. Quidel implementation/tech support has not been very forthcoming with extra help or advice. How did your facility conduct patient correlations at implementation for Troponin I?

I'll email the protocol that we followed. Siemens high-sensitivity Troponin I is our primary method, so we were unable to use it as a comparative method.

Thank you so much, Tara! I appreciate you! My email address is jtruong@stph.org. 

Does anyone else who is a CLIA/Joint Commission have the Triage MeterPro for Troponin I?

Reply
Subgroup Membership is required to post Replies
Join POCT Listserv now
Joy Truong
about 3 years ago
5
Replies
0
Likes
4
Followers
726
Views
Liked By:
Suggested Posts
TopicRepliesLikesViewsParticipantsLast Reply
Requirements for staff performing training
McKenna Chandler
2 days ago
20270
Michael Bishop
1 day ago
Proper Way to Train
Jo Win
3 days ago
60499
Kim Ballister
2 days ago
switching POC glucose from Roche Inform II to Nova Stat Strip
Rosemary Gray
4 days ago
91518
Audrey Fulcher
2 days ago