Coox Comparisons between Multiple GEMs

10 followers
0 Likes

For sites that have multiple GEM blood gas analyzers, what do you use for comparing methhemoglobin, carboxyhemoglobin, oxy and deoxyhemoglobin across the instruments? 
Thanks in advance for your replies!

4 Replies

Lois- we have 5 GEM 5000'S.  We run 5 patient samples and compare it to a GEM that has been designated as the reference.  We use the CAP allowable limits for acceptability.

Thanks Karen! I have 24 GEMs and am looking for something that would be stable enough to run on all instruments across a couple of days. I use PVP for all the other analytes but am looking for something that includes the HGB fractions. CAP has an SO Quality Cross check kit but  it only contains 3 vials. I may also look at using RNA Medical CVC kit.

Our institution has 12 GEM4000 analyzers. We use the manufacturer system evaluators level 1, 2, 3. We also use the CAP TEA (Total allowable error) and compare to 2019: CLIA proposed changes to PT acceptable limits - Westgard our director likes to always have a tighter TEA. Hope this helps. 

More recent link for the newest acceptable PT limits for 2024 (westgaard site):
2024 CLIA Proposed Acceptance Limits for Proficiency Testing - Westgard

Reply
Subgroup Membership is required to post Replies
Join POCT Listserv now
Lois Schultz
over 2 years ago
4
Replies
0
Likes
10
Followers
975
Views
Liked By:
Suggested Posts
TopicRepliesLikesViewsParticipantsLast Reply
Requirements for staff performing training
McKenna Chandler
1 day ago
20259
Michael Bishop
about 10 hours ago
Proper Way to Train
Jo Win
2 days ago
60491
Kim Ballister
about 22 hours ago
switching POC glucose from Roche Inform II to Nova Stat Strip
Rosemary Gray
3 days ago
91507
Audrey Fulcher
1 day ago