QC frequency for Venous Blood Gas on epoc analyzer

7 followers
0 Likes

Hello,

Would anyone be able share their liquid QC schedule for venous blood gas on the epoc analyzer? 
The epoc manual indicates that Federal, State, and Local requirements should be followed. Frequency is not specified.

Thank you !


4 Replies

We use Eurotrol  QC levels 1 and 3 once / month per our IQCP plan.

Thanks Janet! I will set up an IQCP plan for our epoc. 

We use Eurotrol QC levels 1 and 3 every 30 days along with Eurotrol Hct A & B since we do report the H & H from the EPOC as well.

We do Eurotrol 1,3 and Hct A and B with each new shipment of cartridges or monthly if no new cartridges.