iSTAT BhCG comparison to lab

7 followers
0 Likes

For those of you with iSTAT BhCG cartridge and comparing to the laboratory. What are you using as an acceptable criteria when comparing iSTAT BhCG values to your laboratory method?
Thanks
Kim


1 Reply

Our medical director set ours at 25% or 5 mIU/mL.

Reply