I-stat eVAS not updating on Customization Workspace

13 followers
0 Likes

I just updated the latest CLEW (A48) and we are using RALS, everything works fine except for the part where I could not update the e-vas. I was able to upload the latest e-vas
Attachment.
image.png

but when I go back to Customization Workspace, it is still the old e-vas file.
Attachment.
image.png
I have to uncheck the Use eVAS box in order for us to run a QC and just manually compare the ranges from the sheet. I already called the Abbott i-stat customer service and they said they will get back to me. Anybody else having the same issue? 

4 Replies

I had this issue as well. I believe it was a IS security change where I had to regain edit privileges for the DE server. 

Hello Kevin,
You have the wrong eVAS file. The APOC20241371.VAS file is only for the previous CLEW A47.
You need to upload APOC20241373.VAS which is compatible with the new CLEW A48.

Attachment.
eVAS_A47_A48.JPG



Maybe try uploading the file without the file location, like below.  
Attachment.
image.png

I had an issue with actually downloading the evas file and it didn't actually download properly (it said file size was 0.0 kb.  I went back onto the website, re-downloaded the evas file and made sure it was the right file size and then uploaded it into the DE customization workspace and it worked.  

Reply
Subgroup Membership is required to post Replies
Join POCT Listserv now
4
Replies
0
Likes
13
Followers
1360
Views
Liked By:
Suggested Posts
TopicRepliesLikesViewsParticipantsLast Reply
Does a PharmD qualify you as a Technical Consultant
McKenna Chandler
1 day ago
00172
McKenna Chandler
1 day ago
Upgrading Middleware Validations Plan (RALS)
Kristy Saunders
3 days ago
41357
Amanda Casavant
2 days ago
Humidity Monitoring
Jennifer Gagne
5 days ago
100556
Kathleen David
3 days ago