I-stat eVAS not updating on Customization Workspace
13 followers
0 Likes
I just updated the latest CLEW (A48) and we are using RALS, everything works fine except for the part where I could not update the e-vas. I was able to upload the latest e-vas
but when I go back to Customization Workspace, it is still the old e-vas file.
I have to uncheck the Use eVAS box in order for us to run a QC and just manually compare the ranges from the sheet. I already called the Abbott i-stat customer service and they said they will get back to me. Anybody else having the same issue?
but when I go back to Customization Workspace, it is still the old e-vas file.
I have to uncheck the Use eVAS box in order for us to run a QC and just manually compare the ranges from the sheet. I already called the Abbott i-stat customer service and they said they will get back to me. Anybody else having the same issue?
4 Replies
Reply
Subgroup Membership is required to post Replies
Join POCT Listserv now
Suggested Posts
Topic | Replies | Likes | Views | Participants | Last Reply |
---|---|---|---|---|---|
Waived BNP Test | 1 | 0 | 81 | ||
Nova Stat Strip Wi-Fi second generation meters | 0 | 0 | 226 | ||
AMR for BNP ISTAT | 0 | 0 | 129 |
You have the wrong eVAS file. The APOC20241371.VAS file is only for the previous CLEW A47.
You need to upload APOC20241373.VAS which is compatible with the new CLEW A48.