Rapidpoint 500e CVM acceptance criteria
10 followers
0 Likes
We are looking at possibly modifying our acceptance criteria for our Blood Gas analyzers (Rapidpoint 500e) when running the CVMs. For those who use these same analyzers, I would love to know what criteria you use to evaluate your 6 month CVMs. Thank you in advance.
2 Replies
Reply
Subgroup Membership is required to post Replies
Join POCT Listserv now
Suggested Posts
| Topic | Replies | Likes | Views | Participants | Last Reply |
|---|---|---|---|---|---|
| TEG6s for Pediatrics | 3 | 0 | 180 | ||
| Affinion 2 Dropping Numbers from both Operator ID and Specimen ID When Scanning? | 2 | 0 | 194 | ||
| NovaPlus HCG Pregnancy Test Cassettes - Back order? | 1 | 0 | 183 |
We use the criteria that comes with the CVM product itself and it is always in range. I am not sure if you are interfaced through RapidComm or something else, but we just plot the results against each other and it tells us if they're in range or not.