Target times for ACT+ cartridges on GH100
17 followers
0 Likes
For those of you using ACT+ cartridges on the GH100, what are your target times? Mainly for an EP lab performing Ablation procedures, as this is where we will be using these ACT+ cartridges. We currently use the ACT-LR cartridges facility wide, but the providers in EP are wanting to know the actual values during these cases instead of the "out of range high" results they are currently getting, prompting us to add the ACT+ to our test menu. My email is aholland@catawbavalleymc.org if you would prefer to respond via email. Thanks!
1 Reply
Same thing happened at my facility. We started with LR cartridge in the cath labs, but then 1-2 weeks post GO LIVE the MDs requested the higher range (plus cartridge) for EP lab only. After performing in house comparison study to our "old" ACT Plus instrument we ended up sticking to the same target time of >350 seconds for plus cartidge in EP cases.
Reply
Subgroup Membership is required to post Replies
Join POCT Listserv now