i-STAT EG7 cartridge QC Issue

19 followers
0 Likes

Tuesday, we discovered that our lot number of EG7 test cartridges were randomly failing Level 1 POC2 High. We had two lot numbers of QC available, and it was happening with both lot numbers. We are wondering if anyone else is experiencing this issue. The EG7 lot number is N25213, and the Level 1 QC lot numbers are 301190 and 301183.

4 Replies

We have the same lot numbers (Cartridges and QC) , and we haven't had any failures- neither when we validated the lot as new nor when we have validated new iSTAT analyzers.

We are also experiencing the occasional high PO2 but on EG7 lot 2515 with QC lot 301183. Very odd.

Thank you for letting us know. We have had other lots since then without any issues. I do not know what occurred everything was stored at the correct temperature. Who knows what happened between Abbott shipping the cartridges and our receiving them.
Maureen

Please note I do not work on Wednesday.
My work week may be different from your work week. Please do not feel obligated to respond outside of your normal working hours.

Scheduled PTO: 9/29 & 9/30

Maureen J Reynolds MT(ASCP)
Point of Care Coordinator
The Christ Hospital Health Network
2139 Auburn Avenue
Cincinnati, OH 45219
513-585-2809 office
513-585-1434 fax
Maureen.Reynolds@TheChristHospital.com

This thread is interesting, we recently had L3 QC out for TCO2 (reading < reportable range) on the Chem8 and all of the lactates (CG4)  also reading < reportable range.  The same QC lot, different shipment: recovered TCO2 accurately, but not lactate.  We are awaiting a new lot coming from Abbott technical service.  
All of our handling on our side seems consistent; the temperature cards never read greater than 1.  We have electronic monitoring of refrigerator temperatures, and there have been no alarms.



Reply
Subgroup Membership is required to post Replies
Join POCT Listserv now
Maureen Reynolds
11 months ago
4
Replies
0
Likes
19
Followers
1004
Views
Liked By:
Suggested Posts
TopicRepliesLikesViewsParticipantsLast Reply
TEG6s for Pediatrics
Brittany Behrendt
about 23 hours ago
30175
Kristin Cox
about 7 hours ago
Affinion 2 Dropping Numbers from both Operator ID and Specimen ID When Scanning?
Annie Janssen
1 day ago
20188
Lori Lasbury
1 day ago
NovaPlus HCG Pregnancy Test Cassettes - Back order?
Orlando Dominguez
1 day ago
10181
Mi Hui Scales
1 day ago