Miranda Cotter posted a reply on Re: POC Critical result repeating
Saturday, Apr 6, 2024 at 1:57 PM
Said: We require repeats on POC Glucoses because of the high risk of variability that can happen with operator technique. We do require some blood gas repeating based on results, such as if they have 3 or more critical gas results on a ...
Miranda Cotter posted a reply on Re: iSTAT Capillary BGs on peds patients
Saturday, Apr 6, 2024 at 1:46 PM
Said: We switched from the ISTAT blue CG4 to the white CG4 cartridge a couple years back. We ran 10 capillary samples during that validation. We used volunteers for capillary sticking (not too many want to volunteer!). We di...
Cindy Roller posted a reply on Re: Clinitek Status users- 'Insufficient Sample...
Friday, Apr 5, 2024 at 5:34 PM
Said: Hi Melissa - when we start to see an increase in either "borderline" or "insufficient sample" errors, we clean the tray, the insert, and the whole arm (probably not the correct terminology, but I'm sure you know what I'm referring to). ...
Melissa Willcox posted a reply on Re: POC Critical result repeating
Friday, Apr 5, 2024 at 5:02 PM
Said: This is a somewhat complicated issue. Can I ask who is your accreditation agency and if you are specifically referring to POC locations?
Melissa Willcox posted a new Topic
Friday, Apr 5, 2024 at 4:58 PM
Clinitek Status users- 'Insufficient Sample Errors"- what are you seeing
HI again all,
After using the Clinitek Status instruments in multiple locations for use, with a fairly good track record, we have started to see a dramatic increase in the number of times we are seeing the 'Insufficient Sample" error come across necessitating retesting and therefore increased consumption of the Clinitest cassettes- kching$$$, kching$$$.
I do know that when receiving this error in the past, I could associate it with an over filling of the cassette...
After using the Clinitek Status instruments in multiple locations for use, with a fairly good track record, we have started to see a dramatic increase in the number of times we are seeing the 'Insufficient Sample" error come across necessitating retesting and therefore increased consumption of the Clinitest cassettes- kching$$$, kching$$$.
I do know that when receiving this error in the past, I could associate it with an over filling of the cassette...
MOST ACTIVE MEMBERS
James Beck
POCT Coordinator
UPMC
Peggy Mann
Ambulatory POCC & Program Manager
University of Texas Medical Branch
Kathleen David
Associate Director, Point of Care Testing
TriCore Reference Lab
Danyel Olson
POCC
Children's MN
Erika Deaton-Mohney BS, MT(ASCP), CPP
POC and Compliance Coordinator
Bronson
Pet Maniquis
POCC
Providence Medical Center
SHARED LINKS
No Links added yet...
UPCOMING EVENTS
This area will display upcoming events as they approach on the calendar
Add New Module