Said: That is strange that the repeat test on the same machine tested correctly. Maybe I will have my site test that out. Please keep me posted and I will do the same as we investigate. I will let you know if this new lot helps the issue too....
Said: I have had the same issue for 3 cases on multiple devices in a high volume clinic. The repeat test for each case measured at a higher value that fit the expected patient result. These were not due to collection technique. All were run u...
Said: Hello, yes I have the same issue. Lots 0555 and 0558 and several others and multiple analyzers. However they are running the same patient on multiple analyzers (several collections) and results have been 1.3-1.6 percentage off from comp...
Said: So, error E67 indicates an issue with Sample Flow.  Is the alternate control from the vendor and just a different lot number or is it a different vendor's product altogether?  I know that the Chex-Stix is a "prepared" product ...